  fault message(故障信息) possible causes(可能的原因) remedy(补救办法)

  flow too low during measuring (在测量流量太低)

  1:sample gas blocked or leaky(样品气堵塞或泄漏)

  A:clean or replace the blocked parts(hose filter etc) if the problem is still present(清洁或更换堵塞的配件(软管过滤器等)如果问题仍然存在)

  2:pump not running(泵没有运行)

  A:switch the pump on(开关泵)

  B:pump capacity too low(泵的容量太低)

  C:increase the pump capacity as described in section(提高泵的容量中所描述的部分)

  D:pump faulty(泵故障)

  no data for temperature compensation(没有数据进行温度补偿)

  1:temperature compensation not completed successfully(温度补偿没有成功完成)

  2:new components have been loaded(新的组件被加载)

  3:EEPROM has been initialized(EE PROM已初始化)flow too low during autocal(在autocal流量太低)

  A:cal gas path blocked or leaky(校准气路堵塞或泄漏)

  B: clean or replace the blocked parts (hose filter etc) if the problem is still present(清洁或更换堵塞的配件(软管过滤器等)如果问题仍然存在)

  conc of O2 too low during autocal (浓度的氧气在autocal太低)

  1:O2 sensor faulty or old(氧传感器的故障或老化)

  A:replace the O2 sensor as described in section(更换氧传感器中描述的部分)meas value display( 测量值显示)

  2:O2 sensor zero not calibrated(氧传感器的零点未校准)

  calibrate the zero of the o2 sensor as described in section(校准零的氧气传感器中描述的部分)fault at analog output(故障模拟输出)

  A:component could not be initialized when switching on(组件无法初始化时开关)

  B:limits were violated when calibrating the analog section(限制受到侵犯时,校准模拟部分)general fault of all ir channels mv-c1/2/3 1:chopper faulty(斩波器故障)beyond tolerance(一般故障的红外通道mv-c1 / 2 / 3超出公差)

  IR source faulty(红外源错误)fault of addresses for ir channels(故障红外通道地址)

  1:plug-in jumper on detector for detection of components not ok(插件在探测器检测元件不好)

  2:the cable of the detector has no contact(电缆探测器没有接触)

  A:check that the plug is correctly connected to the detector (the plug must latch in twice)(检查插头是否正确连接到探测器(插头锁在两次))

  B:cable of detector faulty(电缆故障探测器)meas value channel 1 beyond tolerance(测量值超出公差1频道)

  analyzer section of first component is faulty(分析仪部分第一组件是错误的)meas value display:*****(测量值显示:*****)

  meas value o2 beyond tolerance(测量氧气超出公差值)

  1:replace the O2 sensor as described in section(更换氧传感器中描述的部分)meas value display:****(测量值显示:* * * *)

  2:motherboard faulty(主板故障)mains voltage beyond tolerance(电源电压超出公差)

  1:mains voltage varies(电源电压变化)

  A:correct such that the mains voltage is within the tolerance values permissible for the analyzer(正确的电源电压范围内允许的公差值分析仪)

  power spply unit faulty(供电单元故障)temp of analyzer beyond tolerance(温度超出公差分析)

  1:ambient temperature too high or too low(环境温度过高或过低)

  A:provide sufficient ventilation or air conditioning(提供充分的通风或空调)

  2:temperature sensor of a detector faulty(温度传感器探测器故障)pressure of amb air beyond tolerance(压力的空气超出公差)

  A:pressure sensor faulty(压力传感器故障)autocal drift beyond tolerance(autocal漂移超出公差)